Heraeus Medical GmbH (Dental Division): Heraeus, PALAMIX uno. Material Number: 66057893.

Medical DevicesSource: FDA (device) · Date: 2026-05-06

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Heraeus Medical GmbH

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Heraeus Medical GmbH (Dental Division)
Product
Heraeus, PALAMIX uno. Material Number: 66057893.
Reason / hazard
Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
Classification
Class III
Status
Ongoing
Distribution / sold at
Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;
Recall initiated
2026-03-25
Code info
Material Number: 66057893; UDI-DI: 04260102133587; Batch: all batches with numbers below 223060;
Product quantity
51,771 units
Voluntary mandated
Voluntary: Firm initiated
City
Wehrheim
Country
Germany
Event id
98608
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.