Abbott Point Of Care Inc.: i-STAT G3+ cartridge; List Number: 03P78-26;
Read the official notice at FDA (device) →
Get alerts for Abbott Point Of Care
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Abbott Point Of Care Inc.
- Product
- i-STAT G3+ cartridge; List Number: 03P78-26;
- Reason / hazard
- Approximately 7.6% of cartridges from specific i-STAT EG7+ (list number 03P76-25), i-STAT EG6+ (list number 03P77-25) and i-STAT G3+ (list number 03P78-26) cartridge lots in the U.S. marketplace may report higher than expected PCO2 and lower than expected pH results due to a manufacturing issue. A falsely high pCO2 together with a falsely low pH makes the blood gas appear as if the patient is in severe respiratory acidosis. If the clinician believes the values are real, it may trigger unnecessary or even harmful interventions such as unnecessary ventilator changes, unnecessary medications, potential harm from overcorrection, or delays in identifying true problems.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of Canada, AUSTRIA, BANGLADESH, BELGIUM, FINLAND, FRANCE, GERMANY, IRELAND, ITALY, LUXEMBOURG, MAURITIUS, NETHERLANDS, NORWAY, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, UNITED KINGDOM.
- Recall initiated
- 2026-04-01
- Code info
- List Number: 03P78-26; UDI-DI: 00054749005673; Lot number: N25221;
- Product quantity
- 40275 units (US only)
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Princeton
- State
- NJ
- Country
- United States
- Event id
- 98635