Intuitive Surgical, Inc.: Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Versio
Read the official notice at FDA (device) →
Get alerts for Intuitive Surgical
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Intuitive Surgical, Inc.
- Product
- Brand Name: Da Vinci X and Da Vinci Xi Product Name: da Vinci X, Xi Surgical System Model/Catalog Number: IS4000; IS4200 Software Version: N/A Product Description: he da Vinci Xi, X Surgical systems each consist of three main components: the Surgeon Console (SC), the Vision Cart (VC), and the Patient Cart (PC). The PC is the operative component of the system, and its primary function is to support the instrument arms and camera arm. Each system have four Universal Surgical Manipulators (USMs) also known as arms, which allow the surgeon to control up to three instruments and an endoscope simultaneously to perform surgical tasks (e.g. cutting, suturing, tissue approximation) in a variety of procedures. As the surgeon moves the hand controls within the console, the PC moves the robotic instruments within the patient in a corresponding manner. The da Vinci system and its software are designed such that instruments motion precisely tracks the motion inputs from the hand controls during surgery. Component: The affected part based on this recall, 380647-xx, USM, is a sub component of da Vinci Surgical Instrument
- Reason / hazard
- Due to manipulator containing screws with in the arm sub-assembly that my be susceptible to breaking.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Norway, Poland, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, and United Kingdom.
- Recall initiated
- 2026-04-02
- Code info
- Lot Code: UDI: 00886874114216 FDA Medical Device Listing Number: D531324 See attachment Affected Product List UDI: 00886874114216 FDA Medical Device Listing Number: D295931 See attachment Affected Product List
- Product quantity
- 2781
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Sunnyvale
- State
- CA
- Country
- United States
- Event id
- 98564