Boston Scientific Corporation: LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
Read the official notice at FDA (device) →
Get alerts for Boston Scientific
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Boston Scientific Corporation
- Product
- LUX-DX II, Arrythmia detector and alarm, Model M302, with SERVER SW LATITUDE DRAGON US, Model 6460, Version 7.5
- Reason / hazard
- For some patients upgraded from a LUX-Dx M301 device to a new Model M302 or M312, the new ICM device is not collecting PVC Burden data (both M302/M312) or monitoring for Bradyarrhythmia and Pause (M312 only). In addition, for these devices that are not being monitored, the LATITUDE Clarity programming screens have discrepant information, indicating that the monitoring is enabled for these features in one location and not enabled in another location.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the country of Belgium.
- Recall initiated
- 2026-03-30
- Code info
- LUX-DX II UDI-DI: 00802526620713; SERVER SW LATITUDE DRAGON US UDI-DI: SERVER SW LATITUDE DRAGON US; Serial Number: 104263 109787
- Product quantity
- 2 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Saint Paul
- State
- MN
- Country
- United States
- Event id
- 98679