Erbe Medical, LLC: ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Sco

Medical DevicesSource: FDA (device) · Date: 2026-05-13

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Erbe Medical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Erbe Medical, LLC
Product
ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208
Reason / hazard
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
Recall initiated
2026-03-31
Code info
REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506, W4464507, W4464512, W4464514, W4464515, W4464516, W4464517, W4465114, W4465115, W4465116, W4465702, W4465709, W4465991, W4466004, W4466266, W4466267, W4466268, W4466269, W4466271, W4466272, W4466273, W4466275, W4466278, W4466279, W4466813, W4466814, W4466815, W4466816, W4467057, W4467058, W4467059, W4467680, W4467681, W4468147, W4468148, W4468273, W4468274, W4468275, WO468776, WO468777, WO468778, WO468779, WO468780, WO469443, WO469444, WO469448, WO469449, WO469450, WO469451, WO469452, WO469933, WO469934, WO469947, WO469948, WO469949, WO469950, WO469951, WO469952, WO470768, WO470774, WO470775, WO470776, WO470777, WO470778, WO470896, WO470897, WO470913, WO470914, WO470915, WO470922, WO471236, WO471237, WO471238, WO471239, WO471240, WO471241, WO471242, WO473987, WO473988, WO473989, WO473991, WO475246, WO475247, WO475248, WO475249; 20325-240/04065655000405/W4466020, W4466021, WO468768, WO468769, WO468770, WO468771, WO469445, WO469446, WO469447, WO469936, WO469937, WO471244, WO471246, WO473170, WO473171, WO473172, WO473173, WO473174, WO473982, WO473983, WO473984, WO473985, WO474409, WO474410, WO474411, WO474412, WO474413, WO474414, WO475243; 20325-206/04065655000436/W4467039, W4467049, W4467678, W4467679, WO468582, WO470769, WO470770, WO470771, WO470772, WO470773, WO471076, WO471077, WO471078, WO471079, WO471390, WO478365, WO478366, WO478367; 20325-207/04065655000689/W4469939, W4469940, W4469941; 20325-248/04065655001006/W4467050, W4467051, W4467055, W4467056, W4467682, WO470898, WO470899, WO470900, WO470901, WO470902, WO470903, WO471252, WO471253, WO471254, WO471255, WO471256; 20325-208/04065655001099/W4467040, W4468271, WO468760
Product quantity
260,962
Voluntary mandated
Voluntary: Firm initiated
City
Tempe
State
AZ
Country
United States
Event id
98674
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.