⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Medline Industries, LP: Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBN

Medical DevicesSource: FDA (device) · Date: 2026-05-13

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Medline Industries

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Medline Industries, LP
Product
Medline Convenience Kits: 1) KIT SURG ONC EXCISION BREAST, Model Number:DYKMBNDL84A; 2) KIT SURG ONC EXPLORATORY LAP, Model Number:DYKMBNDL87A; 3) CT BIOPSY TRAY, Model Number:DYNDH1143C; 4) BIOPSY SPECIALS PACK, Model Number:DYNDH1873; 5) BIOPSY SPECIALS PACK, Model Number:DYNDH1934; 6) BREAST BIOPSY PACK, Model Number:DYNJ17489I; 7) C T BIOPSY TRAY-LF, Model Number:DYNJ17558D; 8) RIGHT HEART BIOPSY TRAY, Model Number:DYNJ19063K; 9) VOR PACK-LF, Model Number:DYNJ48539; 10) VOR PACK-LF, Model Number:DYNJ48539A; 11) IR BIOPSY PACK, Model Number:DYNJ59897B; 12) BREAST PACK, Model Number:DYNJ82998; 13) LIVER BIOPSY KIT, Model Number:DYNJ910654
Reason / hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution.
Recall initiated
2026-01-07
Code info
1) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBQ267; 2) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 22IBU924; 3) DYKMBNDL84A, UDI-DI: 10195327240486(each), 40195327240487(case), Lot Number: 23ABK216; 4) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBK913; 5) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 22LBM302; 6) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23ABU170; 7) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBC364; 8) DYKMBNDL87A, UDI-DI: 10195327240479(each), 40195327240470(case), Lot Number: 23CBL439; 9) DYNDH1143C, UDI-DI: 10193489193336(each), 40193489193337(case), Lot Number: 22GBZ268; 10) DYNDH1873, UDI-DI: 10195327486464(each), 40195327486465(case), Lot Number: 23KBK476; 11) DYNDH1934, UDI-DI: 10195327519728(each), 40195327519729(case), Lot Number: 24FMF173; 12) DYNJ17489I, UDI-DI: 10195327224783(each), 40195327224784(case), Lot Number: 23CBL337; 13) DYNJ17558D, UDI-DI: 10195327504205(each), 40195327504206(case), Lot Number: 23JLB107; 14) DYNJ19063K, UDI-DI: 10888277828025(each), 40888277828026(case), Lot Number: 24DBL079; 15) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LME482; 16) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 20LMF497; 17) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 21JMC430; 18) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22BMH767; 19) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 22EMD675; 20) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23AMC919; 21) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23DMD933; 22) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23GBI270; 23) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23HBR688; 24) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 23KBD033; 25) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24ABL751; 26) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24DBK656; 27) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24EBQ366; 28) DYNJ48539, UDI-DI: 10888277906228(each), 40888277906229(case), Lot Number: 24GBD309; 29) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 24IBT900; 30) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25EBA588; 31) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25HBK043; 32) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25IBA545; 33) DYNJ48539A, UDI-DI: 10198459112133(each), 40198459112134(case), Lot Number: 25JBA908; 34) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 22LBH614; 35) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23EBQ760; 36) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23FBT785; 37) DYNJ59897B, UDI-DI: 10195327043179(each), 40195327043170(case), Lot Number: 23LBA546; 38) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22JMG410; 39) DYNJ82998, UDI-DI: 10195327222987(each), 40195327222988(case), Lot Number: 22LMF004; 40) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25CBM821; 41) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25DBP029; 42) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25EBU142; 43) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25GBO011; 44) DYNJ910654, UDI-DI: 10198459055843(each), 40198459055844(case), Lot Number: 25HBG682
Product quantity
4064 units
Voluntary mandated
Voluntary: Firm initiated
City
Northfield
State
IL
Country
United States
Event id
98642
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.