ARROW INTERNATIONAL, LLC: Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP;
Read the official notice at FDA (device) →
Get alerts for Arrow International
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- ARROW INTERNATIONAL, LLC
- Product
- Cannon II Plus Hemodialysis Catheter, REF: CS-15242-VSP, CS-15282-VSP, CS-15322-VSP, CS-15362-VSP, CS-15552-VSP, CSD-15242-SP, CSD-15282-SP; Cannon II Plus Hemodialysis Catheter with Arrow Simplicity Micro-Introducer, REF: CS-15242-SPM, CS-15282-SPM, CS-15322-SPM, CS-15362-SPM
- Reason / hazard
- 16F dual-valved splittable sheath introducer, the subject of a supplier recall, was included in impacted hemodialysis kits & sets. Sheath introducer may not split as intended, and may result in withdrawal resistance, prolonged procedure time, pain, bleeding, hematoma, tissue injury, vessel wall injury.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution: US nationwide and countries of: AE, AL, AR, AT, AU, BE, BG, BS, BR, BW, CA, CH, CL, CN, CO, CZ, DE, EC, EE, ES, FR, GB, GR, GT, HU, ID, IE, IN, IT, JP, KW, KZ, LB, LT, LU, MN, MZ, NA, NL, NO, NP, NZ, PA, PE, PF, PH, PL, PT, QA, RO, RS, RU, SA, SG, SL, SK, SZ, TH, TN, TR, TT, UA, VN, ZA, ZM, and ZW
- Recall initiated
- 2026-04-10
- Code info
- REF/UDI-DI/Lots: CS-15242-VSP/30801902195273/33F23C0870, 33F23D0857, 33F23F0719, 33F23K0182, 33F24A0265, 33F24C0062, 33F24D1000, 33F24D1039, 33F24E0803, 33F24H0038, 33F24H0261, 33F24K0505, 33F24M0039, 33F24M0218, 33F25A0141, 33F25C0657, 33F25D0125, 33F25E0305, 33F25E0429, 33F25E0430, 33F25E1131, 33F25F0042, 33F25G0070, 33F25H0295, 33F25H0366, 33F25H0674, 33F25J1015; CS-15282-VSP/30801902195518/33F23C0866, 33F23D0854, 33F23E0572, 33F23F0594, 33F23G0540, 33F23J0544, 33F23K0174, 33F23L0950, 33F24B0010, 33F24C0060, 33F24D0993, 33F24D1040, 33F24E0268, 33F24E0809, 33F24F0155, 33F24J0037, 33F24M0007, 33F24M0040, 33F24M0165, 33F24M0217, 33F25A0406, 33F25B0269, 33F25C0021, 33F25C0045, 33F25C0642, 33F25D0013, 33F25D0713, 33F25F0249, 33F25H0296, 33F25J0051; CS-15322-VSP/20801902195184/33F23C0869, 33F23D0862, 33F23F0723, 33F23H0985, 33F24A0272, 33F24C0066, 33F24E0813, 33F24G0170, 33F24G0185, 33F24H0256, 33F24J0040, 33F24L0183, 33F24M0222, 33F25A0426, 33F25C0022, 33F25C0593, 33F25E1082, 33F25G0024, 33F25H0297; CS-15362-VSP/30801902195525/33F23C0875, 33F23G0546, 33F23K0195, 33F24E0186, 33F24H0039, 33F24L0201, 33F24M0235, 33F25C0647, 33F25E0326, 33F25G0058, 33F25J0043; CS-15552-VSP/30801902195198/33F23G0552, 33F23H0987, 33F23L0961, 33F24C0070, 33F24E0820, 33F24H0257, 33F24L0068, 33F24M0227, 33F25C0654, 33F25E0432; CSD-15242-SP/20801902096610/33F24A0303, 33F25C0688; CSD-15282-SP/20801902096634/33F23K1087, 33F24A0297; CS-15242-SPM/10801902193527/33F23D0867, 33F23K0192, 33F24A0288, 33F24E0802, 33F24M0244, 33F25C0669, 33F25E0338, 33F25G0586; CS-15282-SPM/10801902193688/33F23C0872, 33F23D0864, 33F23G0539, 33F23L0949, 33F24A0278, 33F24C0075, 33F24E0808, 33F24H0254, 33F24J0036, 33F24L1166, 33F24M0233, 33F25A0413, 33F25C0666, 33F25H0824; CS-15322-SPM/30801902193538/33F23D0865, 33F23H1010, 33F23L0955, 33F24E0812, 33F25C0682, 33F25E0335; CS-15362-SPM/20801902193548/33F24G0062
- Product quantity
- 34,568
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Morrisville
- State
- NC
- Country
- United States
- Event id
- 98745