ICU Medical, Inc.: Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer S

Medical DevicesSource: FDA (device) · Date: 2026-05-27

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for ICU Medical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
ICU Medical, Inc.
Product
Oncology Kit w/60" (152 cm) Appx 2.2 ml, Smallbore Ext Set w/ChemoLock Port, Clamp, Anti-Siphon Valve, Spiros; ChemoLock; Syringe Transfer Set w/MicroClave, ChemoLock Port, REF: CL3960
Reason / hazard
Oncology sets/closed system transfer devices with extension sets may leak resulting in therapy delayed if noticed and corrected during setup, but if the leak occurs during therapy this may result in therapy interruption with the potential for the patient to receive less infusion solution than intended. Exposure of hazardous substances to the patient, caregiver, or pharmacy technician is possible.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of GA, MA, OH, NC, CA, ND, MI, TX, FL, AR, CT, PA, WA, WI, CO, VT, NY, NH, KY, IA, KS, AL, NM, IL, MO, NJ.
Recall initiated
2026-03-24
Code info
UDI-DI: 10887709083278. Lots: 13919794, 14134777, 14196319
Product quantity
20,506
Voluntary mandated
Voluntary: Firm initiated
City
San Clemente
State
CA
Country
United States
Event id
98622
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.