Novapproach Spine, LLC: OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-

Medical DevicesSource: FDA (device) · Date: 2026-05-27

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Novapproach Spine, LLC
Product
OneLIF Interbody Cage (REF/Description): 010-107-1007/Small 7 Degree 10mm X 7mm, 010-107-1007-2/Small 7 Degree 10mm X 7mm - Oblique, 010-107-1209/Small 7 Degree 12mm X 9mm, 010-107-1411/Small 7 Degree 14mm X 11mm, 010-107-1613/Small 7 Degree 16mm X 13mm, 010-113-1006/Small, 13 Degree 10mm X 6mm, 010-113-1006-2/Small, 13 Degree, 10mm X 6mm - Oblique, 010-113-1208/Small 13 Degree 12mm X 8mm, 010-113-1410/Small 13 Degree 14mm X 10mm, 010-113-1612/Small 13 Degree 16mm X 12mm, 010-119-1407/19 Degree 14mm X 7mm, 010-119-1609/Small 19 Degree 16mm X 9mm, 010-119-1811/Small 19 Degree 18mm X 11mm, 010-125-1606/Small 25 Degree 16mm X 6mm, 010-125-1808/Small 25 Degree 18mm X 8mm, 010-207-1007/Medium 7 Degree 10mm X 7mm, 010-207-1007-2/Medium 7 Degree 10mm X 7mm - Oblique, 010-207-1209/Medium 7 Degree 12mm X 9mm, 010-207-1411/Medium 7 Degree 14mm X 11mm, 010-207-1613/Medium 7 Degree 16mm X 13mm, 010-213-1206/Medium 13 Degree 12mm X 6mm, 010-213-1408/Medium 13 Degree 14mm X 8mm, 010-213-1610/Medium 13 Degree 16mm X 10mm, 010-213-1812/Medium 13 Degree 18mm X 12mm, 010-219-1406/Medium 19 Degree 14mm X 6mm, 010-219-1608/Medium 19 Degree 16mm X 8mm, 010-219-1810/Medium 19 Degree 18mm X 10mm, 010-225-1807/Medium 25 Degree 18mm X 7mm, 010-225-2009/Medium 25 Degree 20mm X 9mm, 010-307-1006/Large 7 Degree,10mm X 6mm, 010-307-1006-2/Large 7 Degree 10mm X 6mm - Oblique, 010-307-1208/ Large 7 Degree 12mm X 8mm, 010-307-1410/Large 7 Degree 14mm X 10mm, 010-307-1612/Large 7 Degree 16mm X 12mm, 010-313-1205/Large 13 Degree 12mm X 5mm, 010-313-1407/Large 13 Degree 14mm X 7mm, 010-313-1609/Large 13 Degree 16mm X 9mm, 010-313-1811/Large 13 Degree 18mm X 11mm, 010-313-2013/Large 13 Degree 20mm X 13mm, 010-319-1606/Large 19 Degree 16mm X 6mm, 010-319-1808/ Large 19 Degree 18mm X 8mm, 010-319-2010/ Large 19 Degree 20mm X 10mm, 010-325-2007/Large 25 Degree 20mm X 7mm, 010-325-2209/ Large 25 Degree 22mm X 9mm, Implant Kits containing OneLIF Interbody Cages, Name/Kit Family: Kit-OneLIF-10 mm (716)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (717)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (718)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (719)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (722)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (723)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (724)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (733)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (734)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (737)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (738)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (746)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (751)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (761)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (762)/Kit-OneLIF-10 mm, Kit-OneLIF-10 mm (787)/Kit-OneLIF-10 mm, Kit-OneLIF- IMPOneLIF Implant Kit (701)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (702)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (703) NL/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (704)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (705)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (706)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (708)/Kit-OneLIF-IMP - OneLIF Implant Kit, Kit-OneLIF-IMP - OneLIF Implant Kit (709)/Kit-OneLIF-IMP - OneLIF Implant Kit,
Reason / hazard
Intervertebral body fusion system straight inserter failed to properly attach to affected cages, which may result in an inability to securely engage the implant with the inserter, which could lead to procedural delay and/or the need to select an alternative implant.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of NY, CT, CA, TX, FL, MI.
Recall initiated
2026-04-23
Code info
REF/UDI-DI/ Cage Lots: 010-107-1007/G07001010710070/ 103003, 105701, 118001, 129301; 010-107-1007-2/G070010107100720/ 114301, 117509, 128703; 010-107-1209/G07001010712090/ 102714, 105702, 117501; 010-107-1411/G07001010714110/ 103019, 105703; 010-107-1613/G07001010716130/ 103020, 105704, 127701; 010-113-1006/G07001011310060/ 103021, 105705, 118002, 129801; 010-113-1006-2/G070010113100620/ 117510, 127716, 114302; 010-113-1208/G07001011312080/ 103022, 105706, 117502, 126902; 010-113-1410/G07001011314100/ 102715, 105707, 127702; 010-113-1612/G07001011316120/ 103023, 105708, 127703; 010-119-1407/G07001011914070/ 103024, 105709, 125101, 128701; 010-119-1609/G07001011916090/ 102716, 105710, 127001; 010-119-1811/G07001011918110/ 103025, 105711, 127003; 010-125-1606/G07001012516060/ 102717, 117901, 127002; 010-125-1808/G07001012518080/ 103005, 117902, 127704; 010-207-1007/G07001020710070/ 102702, 106201, 118003, 125102; 010-207-1007-2/G070010207100720/ 117511, 124106, 127717, 114303; 010-207-1209/G07001020712090/ 102701, 105401, 124101, 127705, 101201; 010-207-1411/G07001020714110/ 105001, 105402, 101203; 010-207-1613/G07001020716130/ 103007, 105403, 124102, 127706; 010-213-1206/G07001021312060/ 105002, 105404, 121202, 124103, 127707, 101204; 010-213-1408/G07001021314080/ 102704, 105405, 121203, 124104, 127708; 010-213-1610/G07001021316100/ 103008, 105406, 130101; 010-213-1812/G07001021318120/ 103009, 105407, 127709; 010-219-1406/G07001021914060/ 102705, 105408, 117503, 119301, 121201, 124105; 010-219-1608/G07001021916080/ 103010, 105409, 121204, 127710; 010-219-1810/G07001021918100/ 103011, 105410, 127711; 010-225-1807/G07001022518070/ 102706, 113701, 117001, 121205; 010-225-2009/G07001022520090/ 103001, 117903, 127004, 127712; 010-307-1006/G07001030710060/ 102707, 106202, 118004, 129302; 010-307-1006-2/G070010307100620/ 117512, 127718, 114304; 010-307-1208/G07001030712080/ 102709, 105601, 117504; 010-307-1410/G07001030714100/ 102710, 105602, 127713; 010-307-1612/G07001030716120/ 103012, 105603, 117505; 010-313-1205/G07001031312050/ 103013, 105604, 117506, 129303; 010-313-1407/G07001031314070/ 102711, 105605, 117507, 124107, 128702, 130102; 010-313-1609/G07001031316090/ 103014, 105606; 010-313-1811/G07001031318110/ 103015, 105607, 127714; 010-313-2013/G07001031320130/ 103016, 105608, 127715; 010-319-1606/G07001031916060/ 102712, 105609, 114905, 117508; 010-319-1808/G07001031918080/ 103017, 105610, 125103; 010-319-2010/G07001031920100/ 103018, 105611, 125104; 010-325-2007/G07001032520070/ 102713, 105612, 121206; 010-325-2209/G07001032522090/ 105613, 121207, 101206, 101207, 101208
Product quantity
1259
Voluntary mandated
Voluntary: Firm initiated
City
Alachua
State
FL
Country
United States
Event id
98835
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.