Fresenius Kabi USA, LLC: LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Fresenius Kabi USA, LLC
- Product
- LVP software of the Ivenix Infusion System (IIS); Product Code LVP-SW-0005.
- Reason / hazard
- Potential software anomaly that can cause the pump to report a false battery-health value (typically 69%) and trigger a battery-depletion alarm that will shut down the pump unless it is reconnected to AC mains power.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution in the states of CA, CO, FL, GA, GA, ID, IL, MD, MI, MN, MO, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA, WI.
- Recall initiated
- 2026-05-06
- Code info
- Product Code LVP-SW-0005. UDI-DI: 00811505030122. Software Version 5.10.2
- Product quantity
- 32 systems
- Voluntary mandated
- Voluntary: Firm initiated
- City
- North Andover
- State
- MA
- Country
- United States
- Event id
- 98929