Baxter Healthcare Corporation: Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with pus
Read the official notice at FDA (device) →
Get alerts for Baxter Healthcare
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Baxter Healthcare Corporation
- Product
- Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with push-on connector on one end (to attach to the Fluid Selector Valve) and a spike or other appropriate connector on the opposite end to attach the source container.
- Reason / hazard
- Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states of Alabama, Arkansas, Arizona, California, Colorado, Florida, Georgia, Hawaii, Iowa, Idaho, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maine, Maryland, Michigan, Minnesota, Missouri, Mississippi, Montana, North Carolina, North Dakota, Nebraska, New Hampshire, New Jersey, New Mexico, Nevada, New York, Ohio, Oklahoma, Oregon, Pennsylvania, South Dakota, Tennessee, Texas, Utah, Virginia, Washington, Wisconsin, and West Virginia. The countries of Taiwan and Uruguay.
- Recall initiated
- 2026-04-22
- Code info
- Lot Code: Lot numbers: 804084, 804088, and 804089 UDI Number: 00085412475806
- Product quantity
- 105675
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Deerfield
- State
- IL
- Country
- United States
- Event id
- 98852