⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Oculus Technologies of Mexico, S.A. de C.V.: PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

Medical DevicesSource: FDA (device) · Date: 2026-06-10

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Oculus Technologies of Mexico, S.A. de C.V.
Product
PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050
Reason / hazard
The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NY, TN, TX, UT, VA.
Recall initiated
2026-04-27
Code info
UDI:00860009268616 Lot: 25G263
Product quantity
5904 units
Voluntary mandated
Voluntary: Firm initiated
City
Zapopan
Country
Mexico
Event id
98959
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.