PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software k
Read the official notice at FDA (device) →
Get alerts for Philips Medical Systems Nederland B.v
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips Azurion 5M20, Model Numbers: 722228 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro Software key
- Reason / hazard
- During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2023-12-07
- Code info
- UDI (01)00884838099234(21)67, (01)00884838099234(21)49, (01)00884838099234(21)10, (01)00884838099234(21)36, (01)00884838099234(21)64, (01)00884838099234(21)110, (01)00884838099234(21)328, (01)00884838099234(21)149, (01)00884838099234(21)213, (01)00884838099234(21)45, (01)00884838099234(21)159, (01)00884838099234(21)18, (01)00884838099234(21)231, (01)00884838099234(21)61, (01)00884838099234(21)204, (01)00884838099234(21)273, (01)00884838099234(21)76, (01)00884838099234(21)150, (01)00884838099234(21)132, (01)00884838099234(21)189, (01)00884838099234(21)17, (01)00884838099234(21)47, (01)00884838099234(21)38, (01)00884838099234(21)87, (01)00884838099234(21)128, (01)00884838099234(21)32, (01)00884838099234(21)23, (01)00884838099234(21)25, (01)00884838099234(21)62, (01)00884838099234(21)181, (01)00884838099234(21)177, (01)00884838099234(21)48, (01)00884838099234(21)153, (01)00884838099234(21)79, (01)00884838099234(21)80, (01)00884838099234(21)157, (01)00884838099234(21)24, (01)00884838099234(21)187, (01)00884838099234(21)26, (01)00884838099234(21)29, (01)00884838099234(21)122, (01)00884838099234(21)127, (01)00884838099234(21)6, (01)00884838099234(21)123, (01)00884838099234(21)124, (01)00884838099234(21)92, (01)00884838099234(21)215, (01)00884838099234(21)91, (01)00884838099234(21)205, (01)00884838099234(21)20, (01)00884838099234(21)37a.
- Product quantity
- 51 units
- City
- Best
- Country
- Netherlands
- Event id
- 98977