ARROW INTERNATIONAL, LLC: Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55
Read the official notice at FDA (device) →
Get alerts for Arrow International
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- ARROW INTERNATIONAL, LLC
- Product
- Description/REF: PI PICC: 3: 6FR x 55CM w CHG/ DLX-35563-HPKC; AGBA PICC: 2L 5.5 FR X 45 CM/ASK-44552-ABKB; AGBA PICC/DELTA KIT: 5.5 FR X 55 CM/ASK-45552-MHBV; AGBA PICC: 1L 4.5FR x 55CM w Biopatch/DLX-45541-HPKB; PICC/DELTA KIT: 2L 5.5 FR X 50 CM/ASK-45052-MC; AGBA PI PICC KIT: 1L 4.5FR 40CM/CDC-44041-HPK1A; AGBA PI PICC KIT 2-LUMEN: 5.5 FR X 40CM/CDC-44052-HPK1A; AGBA PICC/DELTA FG: 2-L 5.5 FR X 40 CM/CDC-44052-VPS2; PI AGBA PICC KIT: 3-L 6 FR X 50 CM/CDC-45063-HPK1A; AGBA PICC/DELTA KIT: 1-L 4.5 FR X 55 CM/CDC-45541-VPS2; AGBA PICC/DELTA FG: 2-L 5.5 FR X 55 CM/CDC-45552-VPS2; PICC/VPS 1L 4.5 FR/ASK-45541-CHS; PI AGBA PICC KIT: 1L 4.5 FR X 55 CM VPS/ASK-45541-UOI2; PICC KIT: 2L 5.5FR 55CM AGBA VPS/ASK-45552-CHS; PI AGBA PICC KIT: 2L 5.5 FR X 55 CM VPS/ASK-45552-UOI2; PICC KIT: 3L 6FR 55CM AGBA/ASK-45563-CHS; PI JACC KIT: 3L 6 FR X 25 CM CG+/ASK-42563-LTAC; PI JACC KIT: 1-L 4.5 FR X 25 CM CG+/CDC-42541-JX1A; PI JACC KIT: 2-L 5.5 FR X 25 CM CG+/CDC-42552-JX1A; PI JACC KIT: 3-L 6 FR X 25 CM CG+/CDC-42563-JX1A; PI JACC KIT: 2-L 5.5 FR X 30 CM CG+/CDC-43052-JX1A; PI JACC KIT: 3-LUMEN: 6 FR X 30 CM CG+/CDC-43063-JX1A; PICC KIT: 2-L 4 FR X 55 CM/CDC-05542-PK1A
- Reason / hazard
- Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
- Recall initiated
- 2026-04-21
- Code info
- REF(Material)/UDI-DI/Lot(Batch): DLX-35563-HPKC/10801902193275/33F25D0520; ASK-44552-ABKB/10801902140798, 10801902216677/33X25K0002, 33S24A0529; ASK-45552-MHBV/10801902193275/33F25K0372, 33F25M0199, 33F26A1534, 33F24L0348; DLX-45541-HPKB/10801902193275, 10801902208849/33F25G0393, 33F25L0055, 33F26A1570, 33F24D0604; ASK-45052-MC/10801902216660/33S24A1242; CDC-44041-HPK1A/10801902193275/33F24K0430; CDC-44052-HPK1A/10801902193275/33F25D0241; CDC-44052-VPS2/10801902193275/33F25E0514; CDC-45063-HPK1A/10801902193275/33F25D0242; CDC-45541-VPS2/50801902120781, 10801902193275/33S24A0211, 33F25C0699; CDC-45552-VPS2/10801902193275/33F25E0704; ASK-45541-CHS/10801902193275/33F26A1422, 33F24L0897; ASK-45541-UOI2/10801902193275/33F24L0991; ASK-45552-CHS/10801902193275/33F26A1395, 33F25G0260; ASK-45552-UOI2/10801902193275/33F24L0431; ASK-45563-CHS/10801902193275/33F25A0199; ASK-42563-LTAC/10801902203271/33F24F0852; CDC-42541-JX1A/10801902193275/33F26A1469, 33F24J0135; CDC-42552-JX1A/10801902193275/33F25D0561, 33F24L0285; CDC-42563-JX1A/10801902193275, 10801902170863/33F25C0702, 33F25D0562, 33F25G0311, 33F25J0437, 33F24D0507; CDC-43052-JX1A/10801902193275/33F25M0062, 33F24J0120; CDC-43063-JX1A/10801902193275, 10801902188332/33F25G0003, 33F26B0805, 33F25D0240; CDC-05542-PK1A/10801902193275, 50801902116715/33F25H0436, 33F25H0576, 33F24D0325
- Product quantity
- 3,481
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Morrisville
- State
- NC
- Country
- United States
- Event id
- 98793