PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost E
Read the official notice at FDA (device) →
Get alerts for Philips Medical Systems Nederland B.v
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Philips MultiDiagnost Eleva with the following Model Numbers: Model # 708032 for MultiDiagnost Eleva; Model # 708034 for MultiDiagnost Eleva with Flat Detector; Model # 708036 for MultiDiagnost Eleva; Model # 708037 for MultiDiagnost Eleva with Flat Detector; Model # 708038 for MultiDiagnost Eleva with Flat Detector.
- Reason / hazard
- Philips Allura Xper and MultiDiagnost-Eleva systems Hand switch button not fully releasing potentially leading to loss of imaging functionality or unintended radiation exposure and additional contrast injection
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US and ROW.
- Recall initiated
- 2026-05-18
- Code info
- All codes
- Voluntary mandated
- FDA Mandated
- City
- Best
- Country
- Netherlands
- Event id
- 99223