Medline Industries, LP: MEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- MEDLINE Pre-connected Urethral Catheterization Tray and Bag, Red Rubber Latex, 15 Fr, Medline Product Number/SKU (REF) DYND10407
- Reason / hazard
- During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.
- Recall initiated
- 2026-05-04
- Code info
- UDI/DI 10080196990740 (each), 40080196990741 (case); ALL LOTS
- Product quantity
- 4304669 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 98994
Free daily digest + "My products" watchlist