⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Staar Surgical AG: Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is des

Medical DevicesSource: FDA (device) · Date: 2026-06-24

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Staar Surgical AG
Product
Product Name: STAAR ICL Calculation Software Software Version: Version 8.00 Product Description: The STAAR ICL Calculation Software is designed as an aide in the calculation of the size and diopter power (with residual refraction) for physician selection of the lens. For toric ICLs, an Implantation Orientation Diagram (IOD) is also produced to provide the physician with pictorial representation of the axis of rotation and alignment. Note: The STAAR ICL Calculation Software resides on STAAR Surgical's e-commerce website in STAAR ICL planning support (Stella), as referenced in the attached recall communication. Component: No
Reason / hazard
Due to a software coding issues for calculating the implant orientation diagram printout for the implantation for toric implantable collamer lenses.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution including in the states and territory of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV and the countries of Austria, Australia, Belgium, Canada, Switzerland, Czech Republic, Germany Spain, Finland, France, Great Britain, Hong Kong, Ireland, Israel, India, Iran, Italy, Jordan, Korea, Kuwait, Kazakhstan, Morocco, Malaysia, Netherlands, Oman, Philippines, Poland, Saudi Arabia, Sweden, Singapore, Thailand, Turkey, Taiwan, Ukraine, USA, Vietnam.
Recall initiated
2026-04-20
Code info
Lot Code: UDI (01)00840311346288(0812)8.00 Software Version 8.00 The software error occurs if a customer generated an IOD for a toric implantation from April 13, 2026 to April 16, 2026 for multiple toric lens reservations for the same eye using the STAAR ICL Calculation Software.
Product quantity
543
Voluntary mandated
Voluntary: Firm initiated
City
Nidau
Country
Switzerland
Event id
99007
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.