Medtronic Perfusion Systems: Octopus Evolution Tissue Stabilizer, Model TS2000
Read the official notice at FDA (device) →
Get alerts for Medtronic Perfusion Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medtronic Perfusion Systems
- Product
- Octopus Evolution Tissue Stabilizer, Model TS2000
- Reason / hazard
- During the manufacturing process, Medtronic personnel identified an assembly issue where the tubing was incorrectly positioned. The short tube was attached to the bottom of the canister, and the long tube was attached to the top, contrary to the intended design.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution.
- Recall initiated
- 2026-05-12
- Code info
- UDI-DI: 00643169668911; Serial Numbers: 0232992409 0233010063 0233010115
- Product quantity
- 134 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Brooklyn Park
- State
- MN
- Country
- United States
- Event id
- 98966