Medline Industries, LP: Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B

Medical DevicesSource: FDA (device) · Date: 2026-07-08

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Medline Industries

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Medline Industries, LP
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B
Reason / hazard
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
Recall initiated
2026-05-28
Code info
UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720
Product quantity
9,272 kits
Voluntary mandated
Voluntary: Firm initiated
City
Northfield
State
IL
Country
United States
Event id
99209
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.