PHILIPS MEDICAL SYSTEMS: Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades
Read the official notice at FDA (device) →
Get alerts for Philips Medical Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- PHILIPS MEDICAL SYSTEMS
- Product
- Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades
- Reason / hazard
- The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide Distribution including Puerto Rico
- Recall initiated
- 2026-06-11
- Code info
- FCO72800812; FCO72800814 728305 728306 DI (01)00884838059474 728307 DI (01)00884838104600 728311 728231 (01)00884838083325 728232
- Product quantity
- 128
- Voluntary mandated
- FDA Mandated
- City
- Cambridge
- State
- MA
- Country
- United States
- Event id
- 99352