PHILIPS MEDICAL SYSTEMS: IQon Spectral CT IQon Spectral CT Upgrades
Read the official notice at FDA (device) →
Get alerts for Philips Medical Systems
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- PHILIPS MEDICAL SYSTEMS
- Product
- IQon Spectral CT IQon Spectral CT Upgrades
- Reason / hazard
- The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide Distribution including Puerto Rico
- Recall initiated
- 2026-06-11
- Code info
- FCO72800812; FCO72800814 728331 728332, DI (01)00884838059542
- Product quantity
- 42
- Voluntary mandated
- FDA Mandated
- City
- Cambridge
- State
- MA
- Country
- United States
- Event id
- 99352