Masimo Corporation: Radius VSM ECG pre-connected Set, REF: 4842, 4695
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Masimo Corporation
- Product
- Radius VSM ECG pre-connected Set, REF: 4842, 4695
- Reason / hazard
- Two devices recalled: 1) Electrocardiogram (ECG) electrode assembly ECG feature may incorrectly detect extreme bradycardia, tachycardia and trigger alarms on normal heart rate patients, which may distract clinicians away from management of true positive alarms, which could delay treatment; 2) Small and medium size blood pressure cuffs may contain rough edges, may result in skin irritation/redness.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of OK, CA , MD , MA , NY , UT , NC , TN , MN , TX , OH , PA , VA and the countries of Italy, Latvia, Iraq, Saudi Arabia, Germany, Spain, Turkey, France, Australia, Malta, Oman, United Arab Emirates, Martinique, Netherlands, Cyprus, Kuwait, Czech Republic, United Kingdom, Croatia, Switzerland, Belgium, Singapore, Romania, Colombia, Hong Kong, Israel, Finland, Poland, New Zealand, Tunisia, Chile, Algeria, Canada, Denmark, Palestinian Territory, Qatar, Greece, Paraguay.
- Recall initiated
- 2026-02-20
- Code info
- REF/UDI-DI/Lot: 4842/10843997011003/25AJH, 25KMA, 25MSC 4695/10843997016015/24N17, 25BBJ, 25DMH, 25FBR, 24KLV, 24N17, 25AGZ, 25BBJ, 25D10, 25DMH, 25FBR, 25GJM, 25KBF
- Product quantity
- 772
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Irvine
- State
- CA
- Country
- United States
- Event id
- 98481