Medtronic MiniMed, Inc.: MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Read the official notice at FDA (device) →
Get alerts for Medtronic MiniMed
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medtronic MiniMed, Inc.
- Product
- MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
- Reason / hazard
- A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
- Recall initiated
- 2026-05-27
- Code info
- UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
- Product quantity
- 778
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northridge
- State
- CA
- Country
- United States
- Event id
- 99207