⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

ResMed Ltd.: Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patien

Medical DevicesSource: FDA (device) · Date: 2026-08-05

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
ResMed Ltd.
Product
Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.
Reason / hazard
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
Classification
Class I
Status
Ongoing
Distribution / sold at
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.
Recall initiated
2026-06-25
Code info
Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070
Product quantity
168,069 devices globally
Voluntary mandated
Voluntary: Firm initiated
City
Bella Vista
State
N/A
Country
Australia
Event id
99304
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.