Boston Scientific Corporation: FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M0
Read the official notice at FDA (device) →
Get alerts for Boston Scientific
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Boston Scientific Corporation
- Product
- FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01
- Reason / hazard
- Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Global Distribution. US Nationwide.
- Recall initiated
- 2026-06-11
- Code info
- UDI-DI: 00191506043131; Lots: CL12695, CL12696, CL12699, CL12700, CL12703, CL12808, CL12816, CL12817, CL12823, CL12834, CL12835, CL12836, CL12853, CL12860, CL12882, CL12883, CL12884, CL12891, CL12907, CL12908, CL12909, CL12919, CL12949, CL12950, CL12969, CL12976, CL12979, CL12980, CL13005, CL13010, CL13011, CL13012, CL13018, CL13032, CL13033, CL13034, CL13037, CL13056, CL13057, CL13058, CL13074, CL13075, CL13076, CL13077, CL13097, CL13098, CL13099, CL13100, CL13104, CL13115, CL13116, CL13117, CL13118, CL13119, CL13135, CL13136, CL13137, CL13138, CL13139, CL13156, CL13157, CL13191, CL13192, CL13193, CL13194, CL13197, CL13208, CL13209, CL13210, CL13215, CL13216, CL13217, CL13224, CL13229, CL13230, CL13257, CL13258, CL13288, CL13289, CL13337, CL13338, CL13365, CL13366, CL13384, CL13398, CL13399, CL13419, CL13420, CL13423, CL13467, CL13469, CL13470, CL13471, CL13472, CL13517, CL13525, CL13526, CL13527, CL13528, CL13535, CL13536, CL13541, CL13550, CL13551, CL13552, CL13553, CL13555, CL13564, CL13565, CL13566, CL13571, CL13574, CL13575, CL13576, CL13598, CL13599, CL13616, CL13637, CL13642, CL13643, CL13644, CL13665, CL13666, CL13667, CL13668, CL13704, CL13705, CL13706, CL13715, CL13716, CL13721, CL13734, CL13735, CL13736, CL13761, CL13762, CL13763, CL13798, CL13799, CL13800, CL13822, CL13825, CL13826, CL13835, CL13836, CL13837, CL13858, CL13859, CL13860, CL13912, CL13913, CL13914, CL13927, CL13928, CL13929, CL13942, CL13944, CL13945, CL13974, CL13975, CL13981, CL13990, CL13991, CL13992, CL14027, CL14028, CL14029, CL14040, CL14041, CL14042, CL14053, CL14054, CL14055, CL14079, CL14080, CL14081, CL14122, CL14123, CL14124, CL14139, CL14140, CL14141, CL14169, CL14172, CL14173, CL14174, CL14188, CL14189, CL14190, CL14227, CL14228, CL14229, CL14250, CL14251, CL14252, CL14288, CL14289, CL14290, CL14347, CL14348, CL14377, CL14378, CL14398, CL14399, CL14400, CL14440, CL14441, CL14442, CL14489, CL14490, CL14492, CL14493, CL14496, CL14502, CL14505, CL14506, CL14520, CL14583, CL14584, CL14585, CL14634, CL14636, CL14637, CL14671, CL14672, CL14673, CL14674, CL14698, CL14699, CL14700, CL14712, CL14713, CL14714, CL14715, CL14751, CL14752, CL14753, CL14771, CL14772, CL14773, CL14774, CL14815, CL14816, CL14817, CL14818, CL14856, CL14857, CL14858, CL14859, CL14872, CL14873, CL14874, CL14930, CL14931, CL14932, CL14936, CL14937, CL14938, CL14970, CL14971, CL14972, CL15006, CL15007, CL15010, CL15011, CL15068, CL15069, CL15070, CL15080, CL15089, CL15110, CL15130, CL15131, CL15132, CL15194, CL15195, CL15196, CL15241, CL15242, CL15243, CL15273, CL15274, CL15275, CL15276, CL15319, CL15320, CL15321, CL15322, CL15359, CL15360, CL15361, CL15362, CL15398, CL15399, CL15400, CL15401, CL15428, CL15429, CL15430, CL15431, CL15451, CL15475, CL15476, CL15477, CL15491, CL15492, CL15493, CL15494, CL15525, CL15540, CL15541, CL15548, CL15549, CL15570, CL15583, CL15584, CL15599, CL15603, CL15617, CL15641, CL15642, CL15718, CL15756, CL15769, CL15770, CL15777
- Product quantity
- 508,151 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Saint Paul
- State
- MN
- Country
- United States
- Event id
- 99252