ICU Medical Inc.: Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous

Medical DevicesSource: FDA (device) · Date: 2026-07-29

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for ICU Medical

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
ICU Medical Inc.
Product
Hotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperatures under gravity flow conditions, List Number L-70
Reason / hazard
ICU Medical became aware of an increased likelihood of leakage in specific Lots of L-70 HOTLINE Fluid Warming Sets due to potentially inadequate solvent application at the tubing to return connector bond.
Classification
Class II
Status
Ongoing
Distribution / sold at
Worldwide - US Nationwide distribution.
Recall initiated
2026-06-24
Code info
UDI-DI: 30695085407007; Lot Numbers: 4293568, 4293569, 4297510, 4297509, 4326087, 4337724, 4337725, 4346349
Product quantity
39,750 units
Voluntary mandated
Voluntary: Firm initiated
City
Minneapolis
State
MN
Country
United States
Event id
99289
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.