Permobil Inc: Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Permobil Inc
- Product
- Permobil M3 Corpus Power Wheelchair with Heavy Duty Shock REF:120031, Model version 3
- Reason / hazard
- This recall has been reported retroactively. The power wheelchair firmware lacks a speed restriction to maintain stability when the seat lift exceeds 50 mm, which may cause the user to fall from the wheelchair or the wheelchair to tip over,
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KY, ME, MI, MN, MO, MS, MT, NC, ND, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY and the country of CAN.
- Recall initiated
- 2024-02-29
- Code info
- All Units operating with firmware version 1.26.2/UDI:(01)17330818345674
- Product quantity
- 140 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Lebanon
- State
- TN
- Country
- United States
- Event id
- 99315