Instrumentation Laboratory: Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system
Read the official notice at FDA (device) →
Get alerts for Instrumentation Laboratory
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Instrumentation Laboratory
- Product
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system
- Reason / hazard
- Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2026-05-08
- Code info
- GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- Product quantity
- 3
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Bedford
- State
- MA
- Country
- United States
- Event id
- 99115