Pride Mobility Products Corp.: Brand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13
Read the official notice at FDA (device) →
Get alerts for Pride Mobility Products
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Pride Mobility Products Corp.
- Product
- Brand Name: Quantum Rehab Product Name: Quantum Rival Model/Catalog Number: Quantum Rival Software Version: 2.0.11, new version: 2.0.13 Product Description: Wheelchair, Powered (ITI), 890.3860 Component: Controller
- Reason / hazard
- If the "Remind Me Later" option is selected for user reminders, reminder pop-up messages may appear while the wheelchair is in use. If this occurs while driving the wheelchair, the wheelchair may perform an unexpected park brake stop, which may result in fall or injury. If this occurs while not driving the wheelchair, the control system may crash and remain in pre-operational mode.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states of AZ, NJ, PA, OK, CO, IN, AR, KS, CT, MD, WI, TX, FL, WA, NY, MO, LA, UT, OR, VA, RI, CA, TN, MN, KY, NV, GA, WV, IL, OH, SC, WY, AL, MI, ME, NC, SD, DE, MA, NE, ND, IA, NH, NM, MT, ID, MS, DC, AK and the countries of Canada and Australia.
- Recall initiated
- 2026-06-25
- Code info
- Lot Code: Quantum Rival 0060650920013*4 K011993 D049085 41875 ITI
- Product quantity
- 1102
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Duryea
- State
- PA
- Country
- United States
- Event id
- 99286
Free daily digest + "My products" watchlist