MUNEVO GMBH: munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control Sys
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- MUNEVO GMBH
- Product
- munevo DRIVE and munevo DRIVE r-net Powered Wheelchair Hands-Free Control Application/Smart Glasses Interface/Wheelchair Head Control System
- Reason / hazard
- A software issue may cause the wheelchair's driving mechanism to fall out of calibration. When this occurs, the device may experience unexpected steering drift or imprecise control while driving the wheelchair. The wheelchair may veer from its intended path, which may cause in low-speed collision, tipping, or injuries to user or bystanders.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of California, Florida, North Carolina, Texas, Utah, and Washington. The countries of United Kingdom, Spain, Italy, Germany, and Switzerland.
- Recall initiated
- 2026-06-19
- Code info
- Affected software versions: munevo DRIVE v1: 10.14.5 - 12.0.4; munevo DRIVE R-net: 10.17.5. - 12.0.4 UDI-DI 04260691380010 Units with affected features enabled: 2QNTGG2820170029, 2QNTGG2820170047, 2QNTGG2820160730, 2QNTGG2820161031, 1QNTGG2519120215, 1QNTGG2818520491, 1QNTGG2819210671, 2QNTGG2820160358, 2QNTGG2820170527, 2QNTGG2820160265, 2QNTGG2820160732, 1QNTGG2819200073, 2QNTGG2820160746, 2QNTGG2820160410, 2QNTGG2820160500, 2QNTGG2820160368, 2QNTGG2820161091, 2QNTGG2820161026, 2QNTGG2820160699
- Product quantity
- 19
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Munich
- Country
- Germany
- Event id
- 99328