⚠ RecallRadar📷 ScanGuidesOfficial U.S. recall notices, restated plainly

Paragonix Technologies Inc.: Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001

Medical DevicesSource: FDA (device) · Date: 2026-08-26

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Paragonix Technologies Inc.
Product
Paragonix Technologies Inc. KidneyVault Portable Renal Perfusion System REF# PRGNX-8008-001
Reason / hazard
Due to a firmware defect, the portable organ transfer device datalogger may lose its configuration, resulting in a blank screen and loss of displayed preservation parameters.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states and territories of AL, AZ, CA, TN, FL, HI, MN, NJ, TX, UT, CO, CT, DC, MD, GA, KY, LA, MA, MI, MO, MS, NC, ND, NE, NM, NV, NY, PA, PR, SC, UT, VA, WI, WV, IN, OH, KY, IL, MD, MN, OR.
Recall initiated
2026-07-20
Code info
All Serial Numbers/UDI: 00850005470342 Software Version: Paragonix App Ver. 6.1.1/6.1.2
Product quantity
1445 systems
Voluntary mandated
Voluntary: Firm initiated
City
Braintree
State
MA
Country
United States
Event id
99476
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.