Siemens Healthcare Diagnostics, Inc.: Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
Read the official notice at FDA (device) →
Get alerts for Siemens Healthcare Diagnostics
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Siemens Healthcare Diagnostics, Inc.
- Product
- Beckman Coulter, Emit 2000 Quinidine Assay REF: OSR4Q229 Contents: R1: 2 x 23 mL;R2 2x13 mL
- Reason / hazard
- Due to unacceptable precision performance exhibited in certain lots of the Emit 2000 Quinidine Assay, which may lead to erroneously elevated quinidine measurements, creating risk of sub-optimal patient treatment, including potential life-threatening cardiac arrhythmias or inadequate antimalarial therapy.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the state of CA and the country of Italy.
- Recall initiated
- 2026-07-13
- Code info
- Lots: 2597, 2627, 2709/UDI: 00842768023293
- Product quantity
- 55 units
- Voluntary mandated
- Voluntary: Firm initiated
- City
- East Walpole
- State
- MA
- Country
- United States
- Event id
- 99446
Free daily digest + "My products" watchlist