Fort Defiance Industries, LLC: Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Fort Defiance Industries, LLC
- Product
- Fort Defiance Industries (FDI), Automated Field Steam Sterilizer REF: P2131
- Reason / hazard
- The Automated Field Steam Sterilizer's water recovery system 120V Amphenol connector may experience an intermittent loss of ground continuity, which could result in an unreliable fault current return path.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of MD, VA, NJ, TN and the countries of Canada, Lithuania, Spain.
- Recall initiated
- 2026-07-08
- Code info
- Lot Code: All serial numbers/ GTIN: (01) 00862992000327
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Loudon
- State
- TN
- Country
- United States
- Event id
- 99371