GE Medical Systems Information Technologies Inc: PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592

Medical DevicesSource: FDA (device) · Date: 2026-09-02

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for GE Medical Systems Information Technologies

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
GE Medical Systems Information Technologies Inc
Product
PORTRAIT COMPUTE AND CORE SERVICES V1.2, Ref# 5700592
Reason / hazard
Software issue may result in a loss of patient monitoring at the Hub and Central Viewer. In addition, all alarm settings could be reset to factory defaults following the update and will not follow the alarm settings previously configured for each patient or care area.
Classification
Class II
Status
Ongoing
Distribution / sold at
US: MI, NY and Australia, Czech Republic, Finland, Germany, Italy, Netherlands, Qatar, Saudi Arabia, Switzerland, United Arab Emirates, United Kingdom.
Recall initiated
2026-07-20
Code info
GTIN 00198953056752, Serial Numbers: SWT25510001TA, SWT26110010TA, SWT26070001TA, SWT26120008TA.
Product quantity
4 units
Voluntary mandated
Voluntary: Firm initiated
City
Waukesha
State
WI
Country
United States
Event id
99536
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.