Medline Industries, LP: Medline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920A
Read the official notice at FDA (device) →
Get alerts for Medline Industries
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Medline Industries, LP
- Product
- Medline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920A
- Reason / hazard
- The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- US Nationwide distribution.
- Recall initiated
- 2026-06-12
- Code info
- 1) DYNDH1920, UDI-DI: 10653160366317(each), 00653160366310(case), Lot Number: 2023111390; 2) DYNDH1920, UDI-DI: 10653160366317(each), 00653160366310(case), Lot Number: 2023121890; 3) DYNDH1920A, UDI-DI: 10653160385943(each), 00653160385946(case), Lot Number: 2025022790
- Product quantity
- 17,750 kits
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Northfield
- State
- IL
- Country
- United States
- Event id
- 99334