Fresenius Medical Care Renal Therapies Group, LLC: 5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C
Read the official notice at FDA (device) →
Get alerts for Fresenius Medical Care Renal Therapies Group
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- 5008X HD/HDF with CLiC Blood Tubing Set Part number 03-5300-3C
- Reason / hazard
- Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
- Recall initiated
- 2026-08-10
- Code info
- Lots 25JR01801 25KR01225 25KR01251 25NR01016 25NR01265 UDI-DI 00840861102532
- Product quantity
- 18,600
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Waltham
- State
- MA
- Country
- United States
- Event id
- 99556