CooperSurgical, Inc.: INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- CooperSurgical, Inc.
- Product
- INCA (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set. Stay-Flex tubing component. 5 units per box. Model Numbers: 44-2707, 44-2709, 44-2710, 44-2712, 44-2715.
- Reason / hazard
- Potential to exhibit a loose connection between the flexible tubing and the nasal cannula, during assembly or while in use, which may result in loss of CPAP pressure, possibly reducing the effectiveness of treatment and lead to decreased oxygenation until connection is restored.
- Classification
- Class I
- Status
- Ongoing
- Distribution / sold at
- Nationwide distribution to CT, FL, GA, IA, IL, IN, MD, MI, MS, NJ, NY, OH, PA, VA. International distribution to Brazil and Peru.
- Recall initiated
- 2026-08-11
- Code info
- 1. Model Number: 44-2707. UDI-DI: (01)20888937009433. Lot Code: 550019722. 2. Model Number: 44-2709. UDI-DI: (01)60888937009448. Lot Codes: 550031617, 550033781, 550031621. 3. Model Number: 44-2710. UDI-DI: (01)20888937009457. Lot Codes: 550031615, 550031616, 550031620, 550034760. 4. Model Number: 44-2712. UDI-DI: (01)20888937009464. Lot Codes: 550031611, 550031614, 550033779, 550031619. 5. Model Number: 44-2715. UDI-DI: (01)20888937009471. Lot Codes: 550031618, 550033780, 550034763.
- Product quantity
- 2,643 boxes (13,215 units)
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Trumbull
- State
- CT
- Country
- United States
- Event id
- 99643
Free daily digest + "My products" watchlist