Bard Peripheral Vascular Inc: EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with softw

Medical DevicesSource: FDA (device) · Date: 2026-10-07

Read the official notice at FDA (device) →

Share on XFacebookRedditEmail

Get alerts for Bard Peripheral Vascular

Guides: Which agency handles which recall (FDA vs USDA vs CPSC vs NHTSA)

The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.

Company / brand
Bard Peripheral Vascular Inc
Product
EnCor EnCompass Breast Biopsy and Tissue Removal System Console (model number: ENCPCONSOLE) and Driver (model number: ENCPDRIVER) with software
Reason / hazard
Breast biopsy and tissue removal system may experience the following errors: 1) Probe calibration and /or handheld driver failure (code HH01-00020009) , 2) Pressure sensor (code E01-0054000E), which prompts the user to restart and re-calibrate; errors may lead to interruption or prolongation of the biopsy procedure, or the need for additional tissue sampling or repeat biopsy.
Classification
Class II
Status
Ongoing
Distribution / sold at
US Nationwide distribution in the states of KS, SC, CA, TX, GA, NC, PA, SC, ID, MI, KY, CT, IL, VA, MA, WI, NY, IN, TN, OH, UT.
Recall initiated
2026-08-20
Code info
Software Versions 1.1.1.10207 and 1.2.0.12133. UDI-DI: 00801741235221 and 00801741234248. Serial Numbers: VTLNPX06, VTLNPX0F, VTLNPX0D, VTLNPX08, VTLNPX0K, VTLNPX0A, VTLNPX0C, VTLNPX0L, VTLNPX0J, VTLNPX04, VTLNPX0H, VTLNPX02, VTLPPX1G, VTLPPX1D, VTLPPX1C, VTLPPX1M, VTLPPX2H, VTLPPX2I, VTLPPX2F, VTLQPX15, VTLQPX1Q, VTLQPX1G, VTLQPX14, VTLQPX0T, VTLQPX1Y, VTLQPX1M, VTLQPX2G, VTLQPX1O, VTLQPX1K, VTLQPX19, VTLQPX2E, VTLQPX1W, VTLQPX10, VTLQPX1Z, VTLQPX24, VTLQPX21, VTLQPX0X, VTLQPX1E, VTLQPX0U, VTLQPX26, VTLQPX2F, VTLQPX25, VTLQPX0Z, VTLQPX18, VTLQPX28, VTLQPX20, VTLQPX27, VTLQPX11, VTLQPX2B, VTLQPX1F, VTLQPX2H, VTLQPX41, VTLQPX3T, VTLQPX43, VTLRPX0R, VTKYPX09, VTKYPX44, VTKYPX46, VTKYPX47, VTKYPX48, VTKYPX49, VTKZPX14, VTKZPX15, VTKZPX20, VTKZPX17, VTLNPX4I, VTLNPX49, VTLNPX4G, VTLNPX4N, VTLNPX4E, VTLNPX4L, VTLNPX44, VTLNPX46, VTLNPX4C, VTLPPX27, VTLPPX28, VTLPPX29, VTLPPX1Z, VTLPPX37, VTLPPX3B, VTLPPX34, VTLPPX38, VTLPPX36, VTLQPX0Q, VTLQPX0B, VTLQPX0E, VTLQPX0P, VTLQPX0F, VTLRPX0X, VTLRPX0Z, VTKYPX07, VTKYPX06, VTKYPX04, VTLNPX4B, VTLNPX42, VTLNPX4F, VTLPPX3D, VTLQPX0M, VTKYPX51
Product quantity
99
Voluntary mandated
Voluntary: Firm initiated
City
Tempe
State
AZ
Country
United States
Event id
99590
Free daily digest + "My products" watchlist

Get the daily digest, and tell us what to watch. Everything is optional except your email.

Free allergen alerts:

Categories:

Privacy: we store only what you enter here, use it only to send recall alerts you asked for, never sell or share it, and you can unsubscribe anytime. Double opt-in: we will email you a confirmation link before sending anything; alerts start only after you confirm. Every email has unsubscribe and manage-watchlist links.