Percussionaire Corporation: Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number:
Read the official notice at FDA (device) →
The text below is copied from the official FDA (device) record. If anything differs, the official notice is correct.
- Company / brand
- Percussionaire Corporation
- Product
- Intrapulmonary Percussive Ventilation, IPV 1, Product Number: S00065; Phasitron 5 Breathing Circuit with Unified Connector, Product Number: P5-UC, used with In-Line Valve Instruction Manual
- Reason / hazard
- Labeling updated warning not to use Intrapulmonary percussive ventilation (IPV) in conjunction with volatile anesthetic delivery systems because IPV may alter controlled delivery of volatile anesthetic agents, resulting in deeper/lighter anesthetic effects that may result in hypertonia/hypotonia, tachycardia/ bradycardia, elevated intracranial pressure, awareness during anesthesia, oversedation
- Classification
- Class II
- Status
- Ongoing
- Distribution / sold at
- Worldwide - US Nationwide distribution including in the states of CA, PA, OK, GA, TX, OH, WI, WA, NY, MO, FL, DE, LA, SC, WV, NJ, NE, KS, NC, AL, CO, VA, OR, AZ, IL, CT, ID, NH, UT, TN, MD, VT, MS, NV, DC, ME, IA, AR, HI, ND, IN, MT, NM, MN, MI, SD, MA, AK, RI and the countries of Canada, Belgium, Chile, France, Germany, Israel, Italy, Latvia, Netherlands, Norway, Oman, Philippines, Portugal, Qatar, Russia, Saudi Arabia, Spain, Sweden, Switzerland, Turkey, United Arab Emirates, United Kingdom, Kuwait, Australia, South Africa.
- Recall initiated
- 2026-08-18
- Code info
- Product Number/UDI-DI: S00065/00849436000679, 00849436000709; P5-UC/00849436000686 In-Line Valve Instruction Manual: P20174 Rev D
- Product quantity
- 61,858
- Voluntary mandated
- Voluntary: Firm initiated
- City
- Sandpoint
- State
- ID
- Country
- United States
- Event id
- 99617